New TopShim run:

version = topshim 2.4.0
date = 2026-11-20
time = 16:21:43.632
command = topshim  convcomp ordmax=4

FINE SHIMMING

Parameters:

probe = Z104275_0310
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 7.24 ppm

Results:


1D Shimming:
Iteration  1: FWHM Before= 2.9 Hz, Target= 0.5 Hz, After= 0.9 Hz
Iteration  2: FWHM Before= 0.9 Hz, Target= 0.8 Hz, After= 0.9 Hz
Iteration  3: FWHM Before= 0.9 Hz, Target= 0.5 Hz, After= 1.3 Hz
Iteration  4: FWHM Before= 1.3 Hz, Target= 0.4 Hz, After= 0.8 Hz
Iteration  5: FWHM Before= 0.8 Hz, Target= 0.3 Hz, After= 0.5 Hz
Iteration  6: FWHM Before= 0.5 Hz, Target= 0.5 Hz, After= 0.6 Hz
Using the best result as the final shim state.

Lineshape quality factor (1D spectrum, spinning) = 1.05

shim changes: 
Z	-64
Z2	-123
Z3	-5
Z4	+56

duration = 3 min 44 sec
completed successfully